You are seeing this message because your Web browser does not support basic Web standards. Find out more about why this message is appearing and what you can do to make your experience on this site better.


ABOUT ARCHIVES
Advanced Search

Welcome   | My Account | E-mail Alerts | Access Rights | Sign In


  Vol. 11 No. 5, Sep-Oct 2009 TABLE OF CONTENTS
  Archives
  •  Online Features
  Original Article
 This Article
 •Full text
 •PDF
 • Reply to article
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Contact me when this article is cited
 Related Content
 •Similar articles in this journal
 Topic Collections
 •Facial Plastic Surgery
 •Cosmetic Surgery/ Procedures
 •Alert me on articles by topic

Comparison of Incision Closures With Subcuticular and Percutaneous Staples

Harley S. Dresner, MD; Peter A. Hilger, MD

Arch Facial Plast Surg. 2009;11(5):320-326.

Objective  Incision closures should yield safe, effective healing with excellent cosmesis. Subcuticular absorbable staples may combine the advantages of subcuticular suturing with the efficiency of percutaneous stapling. This study compares absorbable subcuticular staples with percutaneous metal staples as a means of incision closure in facial rejuvenation surgery.

Methods  Sixteen patients undergoing endoscopic eyebrow-lift and/or rhytidectomy were studied. Each patient had 50% of their temporal and postauricular skin incisions closed with subcuticular staples oversewn with 5-0 plain gut and the remaining 50% closed with percutaneous metal staples. Incisions were evaluated intraoperatively and at regular intervals for 1 year postoperatively. Intraoperative assessments included device handling, bleeding, tension, and cosmesis. Postoperative assessments included incision integrity, inflammation, and cosmesis. Patients were also interviewed regarding incision appearance and comfort.

Results  During the early postoperative period, metal staples produced greater incisional erythema and crusting. Subcuticular staples produced better tissue eversion, less erythema, equivalent if not superior comfort, and shorter office visits. These differences faded over time. The need to properly engage the subcuticular stapler in the dermis was the principal impediment to optimal stapler use.

Conclusions  Subcuticular staples represent a safe, comfortable, and potentially more rapid alternative to percutaneous staples. Modifications of the subcuticular stapler device are required before its full potential can be realized.


Author Affiliations: Department of Otolaryngology, Division of Facial Plastic and Reconstructive Surgery, University of Minnesota, Minneapolis.